FAQ: What is the temperature range for E3640A series programming, readback and meter accuracy specifications?

The E3640A series programming, readback and meter accuracy performance specifications are warranted as long the instrument temperature range is 25 ± 5 °C. There is no temperature coefficient outside of this window.

Applies to: E3640A, E3641A, E3642A, E3643A, E3644A, E3645A, E3646A, E3647A, E3648A, E3649A